SERVOPRO NanoChrome O2 Background |
| |
includes report software, packaging |
| Impurities options |
|
| N2 Range 0-250ppb LDL 0.3ppb |
INJ/HC with 2 valves |
| Ar Range 0-250ppb LDL 0.5ppb |
INJ/HC with 2 valves |
| H2 Range 0-250ppb LDL 0.8ppb |
INJ/HC with 2 valves |
| CO Range 0-250ppb LDL 0.5ppb |
BF/HC with 2 valves |
| CO2 Range 0-250ppb LDL 0.8ppb |
BF/HC with 2 valves |
| CH4 Range 0-250ppb LDL 0.8ppb |
BF/HC with 2 valves |
| CO2/CH4 Range 0-250ppb LDL 0.8ppb |
BF/HC with 2 valves |
| NMHC Range 0-250ppb LDL 0.8ppb |
BF with 1 valve |
| |
|
| Chassis options |
|
| Master Chassis |
Max. 2 channels with 4 valves |
| PC+ Slave chassis |
2 BF/HC channels |
| Master +Slave1 Chassis |
Max. 4 channels with 8 valves |
| Master +Slave 1 +Slave 2 Chassis |
Max. 6 channels with 12 valves |
| Aux.oven |
O2 trap only for N2 measurement |
| |
|
| |
|
| Carrier Gas |
Helium |
| |
|
| |
|
| Power |
115 VAC, 50-60 Hz |
| |
220 VAC, 50-60 Hz |
| |
|
| Gas Connectors (to match accessories fittings) |
1/8" VCR |
| |
|
| Serial port |
RS-232 |
| |
|
| |
|
| Remote Support |
VNC |
| |
Net Support |
| |
|
| Internal Dilution System |
Not Required( > 1ppm) |
| |
Required( One per Chassis) |
| |
|
| |
|
| |
|
| 04416A1 |
SERVOPRO NanoChrome H2 Background |
| |
includes report software, packaging |
| Impurities options |
|
| N2 Range 0-250ppb LDL 0.3ppb |
INJ/HC with 2 valves |
| Ar Range 0-250ppb LDL 0.5ppb |
INJ/HC with 2 valves |
| CO Range 0-250ppb LDL 0.5ppb |
BF/HC with 2 valves |
| CO2 Range 0-250ppb LDL 0.5ppb |
BF/HC with 2 valves |
| CH4 Range 0-250ppb LDL 0.5ppb |
BF/HC with 2 valves |
| CO2/CH4 Range 0-250ppb LDL 0.5ppb |
BF/HC with 2 valves |
| CO2/CH4/NMHC Range 0-250ppb LDL 0.5ppb |
BF/HC/SEL with 3 valves |
| |
|
| Chassis options |
|
| Master Chassis |
Max. 2 channels with 4 valves |
| PC+ Slave chassis |
2 BF/HC channels |
| Master +Slave1 Chassis |
Max. 4 channels with 8 valves |
| Master +Slave 1 +Slave 2 Chassis |
Max. 6 channels with valves |
| Aux.oven |
H2 trap only for N2 measurement |
| |
|
| |
|
| Carrier Gas |
Helium |
| |
|
| |
|
| Power |
115 VAC, 50-60 Hz |
| |
220 VAC, 50-60 Hz |
| |
|
| Gas Connectors (to match accessories fittings) |
1/8" VCR |
| |
|
| Serial port |
RS-232 |
| |
|
| |
|
| Remote Support |
VNC |
| |
Net Support |
| |
|
| Internal Dilution System |
Not Required |
| |
Required ( One per Chassis) |
| |
|
| |
|
| |
|
| |
|
| 04402A1 |
Argon Purity Analyser Base |
| |
includes report software, 2 SS coils tubing, packaging |
| |
|
| Trace (>10 ppm), Pack 1 |
O2 Impurity - Enter range & LDL |
| |
0-10ppm ≤O2 range ≤ 0-100ppm LDL is 1.5% full scale |
| |
H2 Impurity - Enter range & LDL |
| |
0-10ppm ≤H2 range ≤ 0-100ppm LDL is 1.5% full scale |
| |
N2 Impurity - Enter range & LDL |
| |
0-10ppm ≤N2 range ≤ 0-100ppm LDL is 1.5% full scale |
| |
CH4 Impurity - Enter range & LDL |
| |
0-10ppm ≤CH4 range ≤ 0-100ppm LDL is 1.5% full scale |
| |
CO Impurity - Enter range & LDL |
| |
0-10ppm ≤CO range ≤ 0-100ppm LDL is 1.5% full scale |
| |
|
| |
|
| Trace (>10 ppm), Pack 2 |
Same as Pack 1 + |
| |
CO2 Impurity - Enter range & LDL |
| |
0-10ppm ≤CO2 range ≤ 0-100ppm LDL is 1.5% full scale |
| |
|
| Trace Exact (<10 ppm), Pack 1 |
O2 Impurity - Enter range & LDL |
| |
0-1ppm ≤O2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb |
| Min Range 0-1 ppm |
H2 Impurity - Enter range & LDL |
| |
0-1ppm ≤H2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb |
| |
N2 Impurity - Enter range & LDL |
| |
0-1ppm ≤N2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb |
| |
CH4 Impurity - Enter range & LDL |
| |
0-1ppm ≤CH4 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb |
| |
|
| Trace Exact (<10 ppm), Pack 2 |
Same as Pack 1 + |
| Min Range 0-1 ppm |
CO Impurity - Enter range & LDL |
| |
0-1ppm ≤CO range < 0-10ppm LDL is 1.5% full scale,Min.50ppb |
| |
|
| Trace Exact (<10 ppm), Pack 3 |
Same as Pack 2 + |
| |
CO2 Impurity - Enter range & LDL |
| Min Range 0-1 ppm |
0-1ppm ≤CO2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb |
| |
|
| Carrier Gas |
Argon |
| |
|
| Power |
115 VAC, 50-60 Hz |
| |
220 VAC, 50-60 Hz |
| |
|
| Gas Connectors (to match accessories fittings) |
1/8" Swagelok |
| |
1/8" VCR (if 0-1 ppm range) |
| |
|
| Serial port |
Not Required |
| |
RS-232 |
| |
|
| Remote Support |
VNC |
| |
Net Support |
| |
|
| Dual Inlets (sample + span, with switching) |
Not Required |
| |
Required |
| |
|
| Calibration using background gas |
Not required |
| |
Required |
| |
|
| |
|
| |
|
| 04403A1 |
Helium Purity Analyser Base |
| |
includes report software, SS liners, packaging |
| |
|
| Trace (>10 ppm), Pack 1 |
O2 Impurity - Enter range & LDL |
| |
0-10ppm ≤O2 range ≤ 0-100ppm LDL is 1.5% full scale |
| |
H2 Impurity - Enter range & LDL |
| |
0-10ppm ≤H2 range ≤ 0-100ppm LDL is 1.5% full scale |
| |
N2 Impurity - Enter range & LDL |
| |
0-10ppm ≤N2 range ≤ 0-100ppm LDL is 1.5% full scale |
| |
CH4 Impurity - Enter range & LDL |
| |
0-10ppm ≤CH4 range ≤ 0-100ppm LDL is 1.5% full scale |
| |
CO Impurity - Enter range & LDL |
| |
0-10ppm ≤CO range ≤ 0-100ppm LDL is 1.5% full scale |
| |
|
| Trace (>10 ppm), Pack 2 |
Same as Pack 1 + |
| |
Ar Impurity - Enter range & LDL |
| |
0-10ppm ≤Ar range ≤ 0-100ppm LDL is 1.5% full scale |
| |
Ne Impurity - Enter range & LDL |
| |
0-10ppm ≤Ne range ≤ 0-100ppm LDL is 1.5% full scale |
| |
|
| Trace (>10 ppm), Pack 3 |
Same as Pack 2 + |
| |
CO2 Impurity - Enter range & LDL |
| |
0-10ppm ≤CO2 range ≤ 0-100ppm LDL is 1.5% full scale |
| |
|
| Trace Exact (<10 ppm), Pack 1 |
O2 Impurity - Enter range & LDL |
| |
0-1ppm ≤O2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb |
| Min Range 0-1 ppm |
H2 Impurity - Enter range & LDL |
| |
0-1ppm ≤H2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb |
| |
N2 Impurity - Enter range & LDL |
| |
0-1ppm ≤N2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb |
| |
CH4 Impurity - Enter range & LDL |
| |
0-1ppm ≤CH4 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb |
| |
|
| Trace Exact (<10 ppm), Pack 2 |
Same as Pack 1 + |
| |
CO Impurity - Enter range & LDL |
| Min Range 0-1 ppm |
0-1ppm ≤CO range < 0-10ppm LDL is 1.5% full scale,Min.50ppb |
| |
|
| Trace Exact (<10 ppm), Pack 3 |
Same as Pack 2 + |
| Min Range 0-1 ppm |
Ar Impurity - Enter range & LDL |
| |
0-1ppm ≤Ar range < 0-10ppm LDL is 1.5% full scale,Min.50ppb |
| |
Ne Impurity - Enter range & LDL |
| |
0-1ppm ≤Ne range < 0-10ppm LDL is 1.5% full scale,Min.150ppb |
| |
|
| Trace Exact (<10 ppm), Pack 4 |
Same as Pack 3 + |
| |
CO2 Impurity - Enter range & LDL |
| Min Range 0-1 ppm |
0-1ppm ≤CO2 range < 0-10ppm LDL is 1.5% full scale,Min.50ppb |
| |
|
| Carrier Gas |
Helium |
| |
|
| Power |
115 VAC, 50-60 Hz |
| |
220 VAC, 50-60 Hz |
| |
|
| |
|
| Gas Connectors (to match accessories fittings) |
1/8" Swagelok |
| |
1/8" VCR (if 0-1 ppm range) |
| |
|
| Serial port |
Not Required |
| |
RS-232 |
| |
|
| Remote Support |
VNC |
| |
Net Support |
| |
|
| Dual Inlets (sample + span, with switching) |
Not Required |
| |
Required |
| |
|
| Calibration using background gas |
Not required |
| |
Required |